Description
multiCare-in strips
Cholesterol
Description
SAMPLE: capillary blood. Minimum volume: 10 mcl.
HEMATOCRIT RANGE: 35-50%
MEASUREMENT RANGE: 130-400 mg/dl (3.3-10.3 mmol/l)
SENSITIVITY: minimum measurable value: 130 mg/dl
ACCURACY: The mean error versus CHOD-PAP (reference method) is <10%;
y = 0.9599 x + 5.4277 is the linear regression line obtained
ACCURACY. Repeatability, the maximum imprecision is <5%; the mean of the coefficient of variation = 2.14% was obtained from measurements made in the laboratory. Reproducibility, the maximum imprecision is <5%; the mean of the coefficient of variation = 3.95% was obtained from measurements made in the laboratory.
READING TIME: 30-60 seconds
TEMPERATURE OF USE OF THE STRIPS: 10-40 °C
EXPIRY AFTER OPENING: 90 days
EXPIRY: 24 months
FACTOR FOR CALIBRATION MEN WOMEN
FASTING NON-FASTING CALIBRATION FACTOR
Format
Pack of 5 pieces (code 3830261)
Pack of 25 pieces (code 3830260)
SAMPLE: capillary blood. Minimum volume: 10 mcl.
HEMATOCRIT RANGE: 35-50%
MEASUREMENT RANGE: 130-400 mg/dl (3.3-10.3 mmol/l)
SENSITIVITY: minimum measurable value: 130 mg/dl
ACCURACY: The mean error versus CHOD-PAP (reference method) is <10%;
y = 0.9599 x + 5.4277 is the linear regression line obtained
ACCURACY. Repeatability, the maximum imprecision is <5%; the mean of the coefficient of variation = 2.14% was obtained from measurements made in the laboratory. Reproducibility, the maximum imprecision is <5%; the mean of the coefficient of variation = 3.95% was obtained from measurements made in the laboratory.
READING TIME: 30-60 seconds
TEMPERATURE OF USE OF THE STRIPS: 10-40 °C
EXPIRY AFTER OPENING: 90 days
EXPIRY: 24 months
FACTOR FOR CALIBRATION MEN WOMEN
FASTING NON-FASTING CALIBRATION FACTOR
Format
Pack of 5 pieces (code 3830261)
Pack of 25 pieces (code 3830260)