Description
multiCare-in strips
Triglycerides
DescriptionSAMPLE: capillary blood. Minimum volume: 10 microl.
HEMATOCRIT RANGE: 30-50%
MEASUREMENT RANGE: 50-500 mg/dl (0.56-5.6 mmol/l)
SENSITIVITY. Minimum measurable value: 50 mg/dl
ACCURACY. The average error versus GPO (reference method) is <10%;
y = 0.9814 x + 3.8374 is the linear regression line obtained.
PRECISION. Repeatability, mean imprecision is <5%; mean coefficient of variation = 4.66% was obtained from laboratory measurements. Reproducibility, maximum imprecision <5%; mean coefficient of variation = 4.62% was obtained from laboratory measurements.
READING TIME: 20-50 seconds
TEMPERATURE OF USE OF THE STRIPS: 10-40 °C
EXPIRY AFTER OPENING: 90 days
EXPIRY: 24 months