Description
SCREEN CHECK TEST
Candida Screen Test
Description
Rapid qualitative self-diagnosis test, capable of detecting the presence of Candida antigens in vaginal secretion samples collected by swab.
The test is based on the immunochromatographic method. The sample collected by swab is added to the sample pad, through the cap containing the dilution buffer. From here the sample migrates to the conjugate pad, releasing anti-Candida antibodies conjugated to blue colored latex particles. The Candida antigen-antibody conjugate complex formed continues to move up to the nitrocellulose membrane strip. The complex is then captured on the support membrane by the immobilized anti-Candida antibodies, generating a signal line.
If Candida antigens are present in the swab sample, a biological reaction will occur and the result will be positive. A blue line will appear in the test line region (T) on the membrane.
The test includes an integrated control system, represented by the control line (C). The control line confirms that the test has been performed correctly.
How to use
Leave the test cassette at room temperature (15°C to 27°C) before performing the procedure.
Open the test kit only when you are ready to perform the procedure. Do not use the test if the pouch is damaged. Have a watch/stopwatch available.
Make sure the surface on which the procedure will be performed is clean. Wash your hands. The test can be performed at any time of the day.
Remove the device from the sealed bag and place it flat on a flat surface. Hold the device steady with one hand. Use the other hand to gently remove the foil seal from the purple cap. The purple cap contains a liquid.
Remove the stick from the wrapper. Insert the stick about 2 cm into the vagina and rotate for 20 seconds. Gently remove the stick.
Rotate the swab in the liquid in the purple cap for 20 seconds. Gently remove the swab, pressing it against the inside wall of the purple cap. Discard the swab.
Hold the device steady with one hand. Rotate the purple cap counterclockwise to the end point, then rotate it counterclockwise to the starting point. Repeat this step two more times. At the end point, the notches on the purple cap should line up with the indentation on the white base.
Read the results after 10 minutes.
Interpretation of results
POSITIVE: a blue line appears at the control line (C) and a further blue line appears at the test line (T).
The test result indicates that Candida antigens are present in the vaginal secretions. Contact a doctor to confirm the result and start treatment.
NEGATIVE: Only one blue line appears at the control line (C) and no blue line appears at the test line (T).
The test result indicates that Candida antigens are not present in the vaginal secretions. If symptoms or complaints persist, it is advisable to contact a doctor.
INVALID: If neither line appears or only the test line (T) appears, the test result cannot be interpreted. You must perform a new test, following the instructions carefully.
Warnings
For in vitro diagnostic use only.
Do not introduce inside the body.
The test is not reusable.
The test only works if the instructions are followed carefully.
A negative test result does not rule out other infections. Contact your doctor.
The test should not be used during menstruation.
Suitable only for women aged 18 years and over.
Do not use during pregnancy.
Do not use the test within 7 days of taking any medicine for vaginal infections.
Do not use the test within 24 hours of using creams, gels, foams, douches or other intimate hygiene products.
Do not use after the expiration date printed on the foil pouch.
Keep out of reach of children.
Dispose of the device and the swab with normal household waste.
Do not use the stick if it falls to the ground.
Do not use test components if they are broken.
Do not use if liquid leaks from the cap. The amount of liquid inside the cap is critical to ensure proper functioning of the test.
Do not twist the cap before removing the seal or inserting the stick. Twisting the top of the device is irreversible and the test will no longer be valid.
Conservation
Store in a dry place at room temperature (between 2°C and 30°C).
Validity with intact packaging: 24 months.
Format
The kit contains:
- 1 test cassette;
- 1 sterile swab;
- 1 instruction sheet for use;
- 1 translated information sheet.
Rapid qualitative self-diagnosis test, capable of detecting the presence of Candida antigens in vaginal secretion samples collected by swab.
The test is based on the immunochromatographic method. The sample collected by swab is added to the sample pad, through the cap containing the dilution buffer. From here the sample migrates to the conjugate pad, releasing anti-Candida antibodies conjugated to blue colored latex particles. The Candida antigen-antibody conjugate complex formed continues to move up to the nitrocellulose membrane strip. The complex is then captured on the support membrane by the immobilized anti-Candida antibodies, generating a signal line.
If Candida antigens are present in the swab sample, a biological reaction will occur and the result will be positive. A blue line will appear in the test line region (T) on the membrane.
The test includes an integrated control system, represented by the control line (C). The control line confirms that the test has been performed correctly.
How to use
Leave the test cassette at room temperature (15°C to 27°C) before performing the procedure.
Open the test kit only when you are ready to perform the procedure. Do not use the test if the pouch is damaged. Have a watch/stopwatch available.
Make sure the surface on which the procedure will be performed is clean. Wash your hands. The test can be performed at any time of the day.
Remove the device from the sealed bag and place it flat on a flat surface. Hold the device steady with one hand. Use the other hand to gently remove the foil seal from the purple cap. The purple cap contains a liquid.
Remove the stick from the wrapper. Insert the stick about 2 cm into the vagina and rotate for 20 seconds. Gently remove the stick.
Rotate the swab in the liquid in the purple cap for 20 seconds. Gently remove the swab, pressing it against the inside wall of the purple cap. Discard the swab.
Hold the device steady with one hand. Rotate the purple cap counterclockwise to the end point, then rotate it counterclockwise to the starting point. Repeat this step two more times. At the end point, the notches on the purple cap should line up with the indentation on the white base.
Read the results after 10 minutes.
Interpretation of results
POSITIVE: a blue line appears at the control line (C) and a further blue line appears at the test line (T).
The test result indicates that Candida antigens are present in the vaginal secretions. Contact a doctor to confirm the result and start treatment.
NEGATIVE: Only one blue line appears at the control line (C) and no blue line appears at the test line (T).
The test result indicates that Candida antigens are not present in the vaginal secretions. If symptoms or complaints persist, it is advisable to contact a doctor.
INVALID: If neither line appears or only the test line (T) appears, the test result cannot be interpreted. You must perform a new test, following the instructions carefully.
Warnings
For in vitro diagnostic use only.
Do not introduce inside the body.
The test is not reusable.
The test only works if the instructions are followed carefully.
A negative test result does not rule out other infections. Contact your doctor.
The test should not be used during menstruation.
Suitable only for women aged 18 years and over.
Do not use during pregnancy.
Do not use the test within 7 days of taking any medicine for vaginal infections.
Do not use the test within 24 hours of using creams, gels, foams, douches or other intimate hygiene products.
Do not use after the expiration date printed on the foil pouch.
Keep out of reach of children.
Dispose of the device and the swab with normal household waste.
Do not use the stick if it falls to the ground.
Do not use test components if they are broken.
Do not use if liquid leaks from the cap. The amount of liquid inside the cap is critical to ensure proper functioning of the test.
Do not twist the cap before removing the seal or inserting the stick. Twisting the top of the device is irreversible and the test will no longer be valid.
Conservation
Store in a dry place at room temperature (between 2°C and 30°C).
Validity with intact packaging: 24 months.
Format
The kit contains:
- 1 test cassette;
- 1 sterile swab;
- 1 instruction sheet for use;
- 1 translated information sheet.