Description
PiC solution
AIRCube 2
Description
CE Class IIa Medical Device (EU Reg. 2017/745).
Aerosol therapy device. Fast, compact and ideal for the whole family. Equipped with ampoule that guarantees the speed and effectiveness of the therapy. With ampoule holder and handle.
Free from latex and phthalates.
Technical specifications
Rated voltage: approx. 230 V.
Frequency: 50 Hz.
Power consumption: 50 W.
Fuse: T2AH250V.
Maximum pressure: 2.57 ± 0.11 bar.
Operating pressure: 0.98 ± 0.06 bar.
Protection against liquids: IP 21.
Appliance operating conditions: 20 minutes ON - 40 minutes OFF.
MMAD:
Respirable fraction:
Ampoule capacity: maximum 8 ml - minimum 2 ml.
Noise level: 60 dBA @50 cm - 58 dBA @100 cm.
How to use
Indicated for aerosol therapy treatment.
Operating temperature: between -5°C and +40°C.
Operating humidity: 15% to 85% RH.
Operating pressure: between 700 and 1,060 hPa.
Cleaning and maintenance: Remove the mask from the nebulizer. Disconnect the connecting tube from the nebulizer. Disassemble all nebulizer components. Discard any remaining medication. Wash the components in water. Dry by hand or air dry in a clean environment using a soft, clean, lint-free cloth. Assemble the nebulizer and store in a dry place.
Disinfection procedure: Disinfect the ampoule, mouthpiece, masks or nosepiece after the last treatment of the day, according to the following procedure. Use a 70% ethyl alcohol solution:
Step 1. Soak the components in the disinfectant solution for 10 minutes.
Step 2. Remove the components and discard the solution.
Step 3. Rinse components with clean drinking water (60 seconds), shake off excess water and allow to air dry in a clean environment.
Components
Child mask: PVC.
Connecting pipe: PVC.
Ampoule: Polypropylene + Polycarbonate.
Warnings
Never immerse or wet the appliance in water or other liquids. To clean it, use only a soft, clean and dry cloth.
Do not boil the masks.
Never use benzene, thinner or other flammable chemicals for cleaning. This cleaning procedure has been tested for 365 cleaning cycles. For greater hygiene and safety reasons, never use the same accessories for more than one patient. It is recommended to clean and check the integrity of the accessories before and after each use and to replace them in case of damage.
Conservation
Store in a cool, dry place, away from heat sources and sunlight.
Storage temperature: between -25°C and +70°C.
Storage humidity: between 15% and 85% RH.
Storage pressure: between 700 and 1,060 hPa.
Validity for intact package: 12 months.
Format
Aerosol device.
Includes: nebulizer kit, adult mask, pediatric mask, connecting tube, mouthpiece, nosepiece, air filters, integrated nebulizer holder and handle.
Cod. 5803822000010
CE Class IIa Medical Device (EU Reg. 2017/745).
Aerosol therapy device. Fast, compact and ideal for the whole family. Equipped with ampoule that guarantees the speed and effectiveness of the therapy. With ampoule holder and handle.
Free from latex and phthalates.
Technical specifications
Rated voltage: approx. 230 V.
Frequency: 50 Hz.
Power consumption: 50 W.
Fuse: T2AH250V.
Maximum pressure: 2.57 ± 0.11 bar.
Operating pressure: 0.98 ± 0.06 bar.
Protection against liquids: IP 21.
Appliance operating conditions: 20 minutes ON - 40 minutes OFF.
MMAD:
0.1% Albuterol | Fluibron Solution (15 mg/2 ml (of Ambroxol HCl) |
Pulmaxan Suspension (0.5 mg/ml Budesonide) |
|
MMAD | 3.59 ± 0.18 micrometers | 3.85 ± 0.06 micrometers | 4.55 ± 0.16 micrometers |
Respirable fraction:
0.1% Albuterol | Fluibron Solution (15 mg/2 ml (of Ambroxol HCl) |
Pulmaxan Suspension (0.5 mg/ml Budesonide) |
|
Breathable fraction (% <5 micrometers) |
61.2 ± 0.02 | 60.3 ± 0.02 | 50.3 ± 0.03 |
Ampoule capacity: maximum 8 ml - minimum 2 ml.
Noise level: 60 dBA @50 cm - 58 dBA @100 cm.
How to use
Indicated for aerosol therapy treatment.
Operating temperature: between -5°C and +40°C.
Operating humidity: 15% to 85% RH.
Operating pressure: between 700 and 1,060 hPa.
Cleaning and maintenance: Remove the mask from the nebulizer. Disconnect the connecting tube from the nebulizer. Disassemble all nebulizer components. Discard any remaining medication. Wash the components in water. Dry by hand or air dry in a clean environment using a soft, clean, lint-free cloth. Assemble the nebulizer and store in a dry place.
Disinfection procedure: Disinfect the ampoule, mouthpiece, masks or nosepiece after the last treatment of the day, according to the following procedure. Use a 70% ethyl alcohol solution:
Step 1. Soak the components in the disinfectant solution for 10 minutes.
Step 2. Remove the components and discard the solution.
Step 3. Rinse components with clean drinking water (60 seconds), shake off excess water and allow to air dry in a clean environment.
Components
Child mask: PVC.
Connecting pipe: PVC.
Ampoule: Polypropylene + Polycarbonate.
Warnings
Never immerse or wet the appliance in water or other liquids. To clean it, use only a soft, clean and dry cloth.
Do not boil the masks.
Never use benzene, thinner or other flammable chemicals for cleaning. This cleaning procedure has been tested for 365 cleaning cycles. For greater hygiene and safety reasons, never use the same accessories for more than one patient. It is recommended to clean and check the integrity of the accessories before and after each use and to replace them in case of damage.
Conservation
Store in a cool, dry place, away from heat sources and sunlight.
Storage temperature: between -25°C and +70°C.
Storage humidity: between 15% and 85% RH.
Storage pressure: between 700 and 1,060 hPa.
Validity for intact package: 12 months.
Format
Aerosol device.
Includes: nebulizer kit, adult mask, pediatric mask, connecting tube, mouthpiece, nosepiece, air filters, integrated nebulizer holder and handle.
Cod. 5803822000010